Customization: | Available |
---|---|
After-sales Service: | Free Online Technical Support |
Application: | Laboratory Apparatus |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Brief Introduction
F97 fluorescent spectrophotometer is a new generation of high performance molecular luminescence analysis instrument. The product structure is exquisite, has the characteristics of high detection sensitivity, fast scanning speed, wide spectrum measuring range, high dynamic range, fast 3D scanning, and so on. Easily meet the requirements in the field of material research, pharmaceutical analysis, biochemical and clinical testing, water quality analysis and control, food safety testing (dairy products, aquatic products, such as vitamin C, selenium, aflatoxin), and other areas.Specification of F97 Series Fluorescence Spectrophotometer | |
Excitation Source | 150W xenon lamp(Hamamatsu) |
Excitation Wavelength | 200nm~900nm |
Emission Wavelength | 200nm~900nm |
Excitation Slit | F97XP/F97Pro:2nm,5nm,10nm,20nm |
F97:10nm | |
Emission Slit | F97XP/F97Pro:2nm,5nm,10nm,20nm |
F97:10nm | |
Wavelength Accuracy | F97XP:±0.4nm |
F97/ProF97:±1.0nm | |
Wavelength Repeatability | F97XP:≤0.2nm |
F97/F97Pro:≤0.5nm | |
Signal-to-Noise Ratio | F97XP:Raman peak of water(P-P):S/N≥200(10nm Slit) |
F97/F97Pro:Raman peak of water(P-P):S/N≥150(10nm Slit) | |
Limit | F97XP:≤5×10-11 g/ml(Quinine Sulfate Solution) |
F97/F97Pro:≤1×10-10 g/ml(Quinine Sulfate Solution) | |
Linearity | γ≥0.995 |
Peak Repeatability | ≤1.5% |
Stability(10min) | Zero Drift:±0.3 |
Value Limit:±1.5% | |
Wavelength Scan Speed | Multi-speed Level, Maximum at 48000nm/min |
Photometric Quantity Range | 0.00-10000.00 |
Data Transportation | USB2.0 |
Power | 200W |
Power Source | AC 220V/50Hz; 110V/60Hz |
Instrument Dimension | 380×445×310(mm) |
Weight | Net Weight:12kg Gross Weight:14kg |