| Customization: | Available |
|---|---|
| After-sales Service: | Online Technical Support |
| Warranty: | 1 Year |
| Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
|---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
EC80 can simultaneously measures pH, Conductivity, TDS Salinity and temperature without switching probe.
EC80 can switch between °C and °F, available to shut-down automatically.
EC80 includes replaceable pH/Conductivity electrode module.
EC80 has a long life-span battery (>100hrs).
| Specifications | ||||
| Parameters | Range | Resolution | Accuracy | |
| pH | 0.00~14.00 | 0.01 | ±0.01pH | |
| Conductivity | 0uS/cm~20.00mS/cm | 1uS/0.01mS | ±1%FS | |
| ( 0uS/cm~1999uS/cm; 2.00mS/cm ~20.00mS/cm ) | ||||
| TDS | 0mg/L~20.00g/L | 1mg/L/0.01g/L | ±1%FS | |
| ( 0mg/L~1999 mg/L; 2.00g/L ~20.00g/L) | ||||
| Salinity | 0.00~10.00ppt | 0.01ppt | ±1%FS | |
| Temperature | -5.0~60.0ºC | 0.1ºC | ±0.5ºC | |
| 23~140ºF | 0.1 ºF | ±0.9ºF | ||
| General | ||||
| pH calibration | 1, 2 and 3 point calibration. Uses USA and NIST buffer sets. | |||
| Conductivity calibration | 1 point calibration | |||
| Temperature compensation range | -5.0~60.0ºC | |||
| Reference temperature | 15ºC~25ºC | |||
| Temperature coefficient | 0.00~4.00%. Default at 1.91%/ °C | |||
| TDS coefficient | 0.30~1.00. Default at 0.65 | |||
| TDS constant | 0.65 | |||
| Temp sensor | Thermistor, 10k ohms, at 25 °C | |||
| Temp unit | °C and °F | |||