Light chamber design -Special materials to ensure that the light chamber deforms to minimum |
||
Pasing Longge structure with diameter | Roland circle being 400mm | |
Wavelength range | 134-680nm | |
Pixel resolution | 10pm | |
Constant temperature | 32.5 ±0.5°C | |
Concave Grating | ||
Engraved line density | 2400l/mm | |
Primary spectral line dispersion rate | 1.04nm/mm | |
Working Conditions | ||
Working temperature | 15-30 °C | |
Relative humidity | ≤70% | |
Power supply | 220±5V, single-phase 50Hz, grounding resistance <10 | |
The laboratory is expected to have no vibration, dust, strong electromagnetic interference, strong airflow, or corrosive gas. | ||
Excitation light source | ||
High energy plasma spark light source technology | High-energy pre-sparking technology (HEPS) | |
Frequency | 100-1000Hz | |
Excitation Platform | ||
3mm analysis spacing on sample platform | Spray electrode technique | |
Dimension/Weight | ||
H435mm, L900mm, W 600mm | 120 kg | |
Power | ||
Maximum power | 1500 W | |
Standby power | 70 W | |
Detector | ||
High-performance linear array CMOS | ||
Analysis time | ||
30 seconds or less, depending on sample type |