System Functions- Fully automatically, discrete, random STAT function - Online test items: at most 40 testing items- Throughput: 260 tests/hour- Analysis Method: end-point, kinetic, two-point, double -reagents,double-wavelength, multi-standard etc, open to various reagents- Data Processing: memory up to 200,000 patients' data
Sample/Reagents Handling- Sample Position: 40 pcs (Include standard,QC,STAT positions)- Sample volume: 2-50µl, 0.1µl step- Sample Probe Cleaning: automatic internal & external washing- Sample Dilution: automatic/Manual pre-dilution, with dilution ratio up to 1:100- Reagent Position: 40 pcs (With refrigerated function, 2ºC-8ºC) - Reagent Volume: 10-500µl, 1µl step- Reagent Probe: automatic liquid level detect, with collision protection function- Reagent Probe Cleaning: automatic internal & external washing Reaction System- Reaction Cuvette: 90 pcs made of special material- Optical Length of Cuvette: 6mm- Reaction Volume: 200~500µl- Reaction Time: 8-10 minutes- Reaction Temperature: 37ºC±0.1ºC- Mixing System: independent mixing probe
Optical System- Light Source: halogen-tungsten lamp- Wavelength: 300-700nm, 9 wavelengths, precision ±2nm- Absorbance Range: 0~4.0Abs- Resolution: 0.001Abs- Spectrophotometry: rear spectrophotometry QC- Various QC managements, insert QC randomly; save, display & print QC chart; define different QC in advance Calibration- Calibration: linear/nonlinear multi-points calibration- Re-test: retest the sample automatically when the result is out of the linearity range or the sample is not sufficient
Cleaning System- 8-step automatic washing, automatic cuvette dry function- Water Consumption: 4L/hour
Working Condition- Power Supply: ~100-240V, 50/60Hz, 1KVA- Temperature: 10ºC-35ºC- Humidity: ≤90%, no dew
Operation SystemWindows XP or Windows 2000
Input and Output- Input: RS-232 interface / keyboard / computer- Output: multi-format printout
Dimension and Weight- Upper cabinet: 700×650×530mm Net Weight: 62kgs- Bottom cabinet: 700×705×620mm Net Weight: 47kgs
Packing Information- Adopts plywood case- Upper Cabinet: 830×880×732mm G.W.: 95kgs- Bottom Cabinet: 800×880×855mm G.W.: 74kgs
Sample/Reagent Position- 40 pcs sample position- 40 pcs reagent position- Reagents refrigerated function- Primary tube and blood serum cup can be used
Sample/Reagent/Mixing Probe - Probe telflon coating technology- Automatic liquid level detect- Collision protection function- Independent mixing probe- Probe internal/external washing- Clot detection function
Reaction Cuvettes- 90 pcs reusable cuvettes- Automatic blank testing for cuvette- Colorimetry system in the pattern of light splitting behind cuvette- Dismountable, easily replace cuvette
8-step auto cleaning System- Automatic 8-step washing- Automatic cuvette dry function- Hi-tech washing station to ensure accurate test results- With low water consumption: 4L/H
- Automatic washing cuvette when start up & shut down the instrument- Test order setting program is available to avoid carry-over- Automatic detect reagent volume during every startup of the instrument- Show reagent plate temperature during every startup of the instrument
Automatic Cuvette blank testing - Automatic wash selected cuvettes - Real-time online monitoring running status of cuvette -
- Real-time online monitoring running status of sample tray, reagent tray cuvette- Real-time online monitoring reagents residual volume- Real-time online monitoring waste volume
Automatic dilute and re-test when absorbance range is overrange -Automatic dilute and re-test when linear limit is overrange -Three modes (no dilute, autodilute and handwork dilute) are available -Freely set the autodilute ratio and dilute water position -
- Freely set the autodilute ratio when re-testing- "No dilute" re-testing mode is available